Biological agents are classified into four risk groups (RG1–RG4) based on their ability to cause disease and availability of treatment. Most clinical laboratories work at Biosafety Level 2 (BSL-2), handling RG2 agents such as hepatitis B virus, HIV, and Salmonella. BSL-2 requires: handwashing facilities, biohazard signage, restricted access, PPE (gloves, gown, eye protection), and biosafety cabinets for aerosol-generating procedures.
|
✓
Good practice The section head reviews the risk assessment: influenza A and B are Risk Group 2 organisms. Open manipulation of respiratory samples requires a Class II biosafety cabinet. Inactivation of samples prior to PCR extraction eliminates infectious risk for downstream steps. The SOP specifies: Class II BSC for receipt and initial processing; standard laboratory safety measures after inactivation. This is proportionate to the risk.
|
✗
Poor practice The section head specifies Level 3 containment measures for all steps including post-inactivation PCR, because influenza can cause severe disease. This requires respirator use in a fully pressurised negative-pressure suite. Staff find the equipment difficult to work in, compliance deteriorates, and a spillage occurs when a scientist struggles to manipulate samples in unfamiliar equipment. Over-specification created a new hazard.
|