Before results are released, they must be validated — checked for clinical plausibility, delta check (consistency with previous results), and internal consistency within the request. Only authorised personnel may release results. Authorisation criteria must be documented. Autoverification (automated result release) is permitted under ISO 15189 provided criteria are documented, validated, and monitored.
|
✓
Good practice On discovering this, the quality manager immediately revises LIS permissions: only scientists with documented competency in result interpretation are assigned authorisation rights. The HCSW’s permission is removed. A nonconformity is raised for the result released without authorisation, and the result is reviewed retrospectively by an authorised scientist.
|
✗
Poor practice The practice continues because removing permissions would slow down urgent result release. Over the following months, several results are released by staff without the competency to assess whether the result is clinically plausible, whether a delta check flag needs review, or whether a critical value notification is required.
|