Patient data in the LIS is subject to data protection law. Access must be role-based and documented. Passwords must meet complexity requirements and be changed at defined intervals. The laboratory must have a documented information security policy. Unauthorised access must be treated as a reportable breach. Staff must not share login credentials under any circumstances.
|
✓
Good practice The section head, reviewing results before release, checks the haemoglobin against the printed analyser output. The discrepancy is immediately visible — the LIS shows 14.2, the printout shows 142. The error is corrected before release, documented in the transcription error log, and the root cause addressed: all manually transcribed results now require a second scientist to verify against the original printout before release.
|
✗
Poor practice The result is released as 14.2 g/L — a value that falls within the laboratory’s female reference interval for haemoglobin (115 to 165 g/L for this lab). The clinician sees a normal result and takes no action. The patient’s true haemoglobin of 142 g/L is also normal — so in this case no harm results. But the process failure is invisible because the error happened to produce a plausible value.
|