Internal QC (IQC) monitors the ongoing performance of examination processes. QC materials must be treated like patient samples. QC must be run before patient testing begins and at regular intervals during the run. Levey-Jennings charts must be maintained and reviewed. QC targets and limits must be established by the laboratory, not taken directly from the manufacturer.
|
✓
Good practice The quality manager reviews the evidence: single-level QC detects bias at the control concentration but cannot detect changes in imprecision that may affect results at other concentrations — particularly at medically critical concentrations above or below the reference interval. She implements two-level QC — one within the reference interval, one in the medically critical zone — run at the start of every analytical run. The additional cost is minimal; the detection capability is substantially improved.
|
✗
Poor practice The single-level QC continues because the laboratory manager says it has always worked and adding a second level would be expensive. During an accreditation assessment, the assessor reviews the IQC strategy and asks how the laboratory detects analytical errors at concentrations outside the normal range. The answer is that it cannot. This is a finding — the IQC strategy is insufficient for the claims of clinical performance.
|