Westgard rules are decision criteria for interpreting QC data. Key rules: 1₂s warning — one result exceeds ±2SD (investigate); 1₃s rejection — one result exceeds ±3SD (reject the run); 2₂s rejection — two consecutive results exceed the same ±2SD (reject); R₄s rejection — range between two consecutive results exceeds 4SD (reject); 4₁s rejection — four consecutive results exceed the same ±1SD (reject). Running multiple QC levels increases error detection.
|
✓
Good practice The analyst recognises the 1-3s rule as a warning for random error. She prepares a fresh aliquot of QC material and re-runs. The fresh aliquot is within range. She investigates the original vial — it shows signs of evaporation from incomplete sealing. Root cause: inadequate QC material storage. The event is documented and patient results are released.
|
✗
Poor practice The analyst re-runs the same vial until it passes. It passes on the third run. She releases patient results with no investigation and no documentation. The underlying storage issue persists. The next morning, QC again shows a 1-3s violation. The pattern of re-running until passing accumulates over weeks — eventually masking a genuine systematic error on the analyser.
|